Percutaneous Removal of a Broken and Embolized Transvenous Chemotherapy Port Catheter in the Left Pulmonary Artery by Using a Snare-loop Catheter
PDF
Cite
Share
Request
Case Report
VOLUME: 55 ISSUE: 1
P: 70 - 73
March 2017

Percutaneous Removal of a Broken and Embolized Transvenous Chemotherapy Port Catheter in the Left Pulmonary Artery by Using a Snare-loop Catheter

Med Bull Haseki 2017;55(1):70-73
1. Haseki Training And Research Hospital, Clinic Of Cardiology, Istanbul, Turkey
No information available.
No information available
Received Date: 10.06.2016
Accepted Date: 10.07.2016
PDF
Cite
Share
Request

ABSTRACT

Totally subcutaneous intravascular portacath provide safe and reliable vascular access and is widely utilized for venous access for long-term parenteral administration of medications. Catheter fracture and/or embolization of the catheter fragment in to the heart and or pulmonary artery is a rare and potentially serious complication. When it occurs, a prompt surgical or percutaneous extraction of the embolized foreign body is necessary. We present an asymptomatic case of metastatic colon adenocarcinoma in a patient who had fragmentation of catheter from the connection of the port and migration to left pulmonary arteries. We successfully removed the 10-cm long and 6-Fr diameter fractured catheter segment from the left pulmonary artery via the right femoral vein using a snare catheter with triple loop without complication. Post-procedure course was uneventful and the nature of the vascular access made early ambulation possible. Compared to surgery, percutaneous approach is a less invasive, safe, reliable and effective technique. Therefore, we suggest that percutaneous transcatheter technique for retrieval of embolized or broken catheter fragments should be considered as the first-choice treatment.

Keywords:
Endovascular, portcath, left pulmonary artery, snare catheter

Introduction

Totally subcutaneous intravascular portacath provides safe and reliable vascular access and is widely utilized for venous access for long-term parenteral administration of medications, such as chemotherapy, in patients with malignancy (1-4). Catheter fracture and or embolization of the catheter fragment in to the heart and or pulmonary artery is a rare and potentially serious complication. When it occurs, prompt extraction of the embolized foreign body is necessary (1,3,5,6). While surgery was performed in the past, currently, endovascular retrieval of embolized catheter fragment is mostly preferred and can be performed easily, safely, and successfully (3,7-11).

We present an asymptomatic patient who had fragmentation of catheter from the connection of the port and migration to left pulmonary arteries. Endovascular removal of the catheter fragment from the left pulmonary artery via the right femoral vein using a snare catheter with triple loop was successfully performed without any complication.

Case

A 34-year-old male with metastatic colon adenocarcinoma diagnosed in July 2015 was receiving monthly chemotherapy through a central venous port implanted into his right subclavian area about three months ago. On December 25, 2015, he experienced local pain during fluid injection in the right subclavian area and impossible fluid injection or blood aspiration via venous port. Chest X-ray revealed migration of the proximal catheter fragment into the left pulmonary artery (Figure 1). Echocardiography showed preserved left ventricular function with ejection fraction of about 55% to 60%. No regional wall motion abnormality was observed. Linear shadow in the left pulmonary artery was seen on echocardiography, representing the retained fragment. After the patient was evaluated by a cardiovascular surgeon and interventional cardiologist; an emergency percutaneous removal of the catheter was planned. Thus, he was referred to the interventional cardiology clinic for minimally-invasive percutaneous removal. Therefore, the patient was admitted to our hospital for retrieval of the embolized catheter fragment on the same day. On admission, his vital signs were stable. On the physical examination, there were subcutaneous swelling and erythema in the right subclavian area, and the other physical examination findings were within the normal limits. The patient and his family were informed in detail about the possible consequences of the presence of a foreign material in the left pulmonary artery. After informed consent was obtained, the patient was taken to the cardiac catheterization laboratory. The procedure was performed under local anesthesia and fluoroscopic guidance. A 6-Fr pigtail catheter was advanced to the left pulmonary artery through the right femoral vein using an 8-Fr sheath. Then, the pigtail catheter was exchanged with an 8-Fr guiding catheter. Approximately 10-cm long and 6-Fr diameter fractured catheter segment that was lodged in the left pulmonary artery was successfully grasped by using a 25 mm snare with triple loops (ev3™Amplatz Goose Neck snare Kit, A111044) and pulled into the right femoral vein along with the sheath and was externalized (Figure 2, 3). We administered intravenous unfractionated heparin at a dose of 70 U/kg (up to a maximum 5000 U) during the procedure. After the activated partial thromboplastin time was maintained at <50 seconds, we removed the femoral vein sheath and hemostasis was achieved through manual pressure. Post-procedure course was uneventful and the nature of vascular access made early ambulation possible.

Discussion

A broken and migrated peripherally inserted central venous port in the right subclavian vein was successfully retrieved using a snare with triple loops.

Port-a-Cath is a totally implantable venous access device in which a conventional central venous catheter is attached to a subcutaneous injection port usually on the chest wall. It has been reported that application of port catheters is safe. Complications after catheter insertion include embolization, infection, venous thrombosis, occlusion of the catheters, venous perforation, atrial perforation, arrhythmias, flebitis, leakage, dislodgement, subintimal entrapment, and fracture and/or migration of the catheters (11-15). Port-a-Catheter fracture with or without embolization is a serious and rare complication in adult patients. The estimated incidence is between 0.2% and 1.1% of all insertions, and the reported mortality rate is approximately 60% (1,5,11,16,17). The present case suffered embolization of a broken catheter material into the left pulmonary artery. The intravascular fragment lodges  more distally within the pulmonary artery especially in the left pulmonary artery with the risk of causing a pulmonary infarction whereas mostly becomes lodged within the right heart (1). Although the exact mechanism of portacath fructure is unknown, the impingement of the port catheter between the first rib and the clavicle by surrounding musculoskeletal system was thought by many authors as a possible cause, known as “pinch off syndrome” or thoracic inlet syndrome. (1,16-20). It was also supposed that factors such as venous flow, negative inspiratory pressure in the thoracic cavity, changes in thoracic pressure with coughing and vomiting, vigorous movement of the upper arms, neck flexion, were thought to lead to migration of fractured material (21,22). In this case, we could not speculate on the cause of migration in our patient.

The patient in our case report complained of local pain, subcutaneous swelling and erythema in the right subclavian area. Most of the patients with the fractured catheter embolization remain asymptomatic. Besides that, they may complain of palpitation, cough, dyspnea, thoracic pain, or local swelling and erythema due to serious complications such as infection, pulmonary embolism, arrhythmia, cardiac perforation, ventricular tachydysrhythmia, cardiac arrest, and endocarditis. The first sign of catheter embolization mostly is catheter malfunction precluding fluid injection or blood aspiration and or local pain and subcutaneous swelling at the injection site as seen in our case (2,16,18,22). Therefore, prompt removal of the catheter fragments should be done.

While surgery was the only choice in the treatment of broken and migrated catheters in the past, currently, percutaneous transcatheter retrieval of the portacath fractured fragment by interventional endovascular techniques is the most common technique and successfully performed with much lower morbidity and mortality rates when compared with surgery in adult patients even in neonates  (1,5,6,14,23-25). The reported success rates of percutaneous retrieval of fractured fragment in the literature are between 71% and 100% (5,17,25).

Although, there are several available tools for retrieving intravascular foreign bodies from the vascular system, such as endovascular forceps, pigtail catheter, ablation catheter and retrieval baskets, snare-loop catheter is the most commonly used devise for intravascular catheter fragment retrieval (1,5,6,23). Additionally, the most often preferred access route is the right common femoral vein due to advantages such as convenience in handling material, access to main sites of venous foreign body migration, the possibility of using larger-calibre sheaths, easier puncture procedure, and safety of effective compression after the end of the procedure (17).

In our case, the fragment was successfully removed from the left pulmonary artery using snare with triple loops via the right femoral vein by endovascular technique under local anesthesia.

Catheter fraction and migration into the left pulmonary artery is a rare and potentially severe life-threatening complication with high mortality rate. Therefore, retrieval of embolized fractured catheter material should be considered. Since percutaneous transcatheter retrieval using loop-snare is a less invasive, safe, reliable and effective technique, which avoids open surgery and surgery-related complications, we suggest that percutaneous transcatheter technique for retrieval of embolized or broken catheter fragments should be considered as the first-choice treatment.

Ethics

Peer-review: Internally peer-reviewed.

Authorship Contributions

Surgical and Medical Practices: Mehmet Mustafa Can, Muhsin Kalyoncuoğlu. Concept: Semi Öztürk. Design: Muhsin Kalyoncuoğlu, Gündüz Durmuş. Data Collection or Processing: Gündüz Durmuş. Analysis or Interpretation: Mehmet Mustafa Can, Mustafa Sarı. Literature Search: Semi Öztürk. Writing: Muhsin Kalyoncuoğlu.
Conflict of Interest: No conflict of interest was declared by the authors.
Financial Disclosure: The authors declared that this study received no financial support.