Failure of Constrained Total Hip Arthroplasty with Intact Reinforcement Ring

Med Bull Haseki 2019;57(4):445-447
1. University of Health Sciences, Haseki Training and Research Hospital, Clinic of Orthopedics and Traumatology, İstanbul, Turkey
2. Beytepe Murat Erdi Eker State Hospital, Clinic of Orthopedics and Traumatology, Ankara, Turkey
No information available.
No information available
Received Date: 27.12.2018
Accepted Date: 17.06.2019
Publish Date: 10.12.2019
PDF
Cite
Share
Request

ABSTRACT

One of the most common complications after total hip arthroplasty (THA) is dislocation. To avoid this unpleasant complication, we use constrained systems especially in patients at high risk of dislocation. We report two cases of failure of constrained hip system with intact reinforcement ring. We used Trilogy® revision cup in both patients. One of the patients refused surgery. Other patient underwent surgery and we observed abrasion of constrained liner because of impingement. Malrotation of the constrained liner appears to cause impingement. When using constrained liners, caution should be exercised against the risk of impingement. Proper assessment of the constrained liner is mandatory. Otherwise, abrasion of the liner is not a surprise. Instructions of the manufacturer should be followed. The fact that such a small misplacement has such great consequences can also be accepted as a disadvantage of the product. It may be difficult to determine the proper rotation of the polyethylene liner during surgery. Therefore, it would be appropriate to have a wider range of correct polyethylene placement angles.

Keywords:
Hip, prothesis, dislocation, constrained

Introduction

One of the most common complications after total hip arthroplasty (THA) is dislocation. The dislocation rate after primary THA has been reported to be 1-7% (1,2). Dislocation after revision hip arthroplasty has been reported in up to 30% of patients (1-3). The constrained liners are indicated as a component of a total hip prosthesis in primary or revision THA for patients who are at high risk for hip dislocation and all other options have been considered. We report two cases of failure of constrained hip system with intact reinforcement ring. We used Trilogy® cup (Zimmer, USA) in both patients. Dislocation rates with this type of acetabular component have been reported to be 9-33% in different series (4,5). Dislocation of constrained hip with intact reinforcement ring is a rare condition. A few published reports of this kind of dislocations exist but we were unable to find any other published reports about Trilogy® constrained cup failure with intact reinforcement ring.

Case Reports

Case 1

A 72-year-old female patient underwent THA in 2010. One year later, she underwent two-stage revision surgery because of infection. In 2014, she presented to our clinic with infected total hip prosthesis and treated with two-stage revision surgery again. In late 2015 she was admitted with periprosthetic distal femur intraartricular fracture. She underwent total femur prosthesis and constrained hip arthroplasty. Five months later, she presented with dislocated hip with intact reinforcement ring (Figure 1).

At revision surgery, the retaining fingers of the polyethylene liner were found at the 2-8 o’clock position and the inferior finger was disrupted because of impingement of the femoral head (Figure 2). New constrained liner was adapted to 1-7 o’clock position which is suggested by the manufacturer.

Case 2

A 79-year-old female patient underwent THA in 2011. Two years later, the acetabular component was revised because of multiple dislocations. In 2015, she applied to our clinic because of hip dislocation with failure of constrained liners. The reinforcement ring was disrupted. Total revision surgery was performed with Trilogy® constrained acetabular liner (Zimmer, USA). Two months later, she presented to our outpatient clinic with the complaint of hip pain. Dislocation of the hip without any disruption of the reinforcement ring was determined on X-rays (Figure 3). We offered revision surgery, but patient preferred to be treated in another clinic.

Discussion

The first case is a patient who was hospitalized for five months because of infection and multiple surgeries. After elimination of infection, revision surgery was performed and the patient returned daily life. The constrained acetabular liner was the source of confidence about stability of the hip. However, unexpectedly, the patient presented with dislocation. There is a risk of dislocation even if the constrained liner is correctly placed. Thoms and Marwin (6) reported a patient with two dislocations of a tripolar constrained acetabular liner caused by impingement. Tufescu and Dust (7) reported two cases of failed constrained THA caused by skirted femoral stem. In our case, we observed impingement because of constrained liner malrotation. The inferior finger of the constrained liner was disrupted. The liner was at the 2-8 o’clock position. A new liner was placed at the 1-7 o’clock position and we gained more range of motion without impingement.

Conclusion

The Trilogy® constrained acetabular liner is a good choice for patients who have high risk of dislocation. Even with constrained acetabular liner, dislocation of the hip is not impossible. When using constrained liners, caution should be exercised against the risk of impingement. Proper assessment of the constrained liner is mandatory. Otherwise, disruption of the liner is not a surprise. Instructions of the manufacturer should be followed. It can also be accepted as a disadvantage of the product that such a small misplacement has such great consequences. It may be difficult to determine the proper rotation of the polyethylene liner during surgery. Therefore, it would be appropriate to have a wider range of correct polyethylene placement angles. Even if there is adequate stability in the operation, we must inform the patient about the allowed range of motion after the operation.

References

1
Soong M, Rubash HE, Macaulay W. Dislocation after total hip arthroplasty. J Am Acad Orthop Surg 2004;12:314
2
Paterno SA, Lachiewicz PF, Delley SS. The influence of patient-related factors and the position of the acetabular component on the rate of dislocation after total hip replacement. J Bone Joint Surg 1997;79-A:1202.
3
Su EP, Pellicci PM. The role of constrained liners in total hip arthroplasty. Clin Orthop 2004;420:122-9.
4
Lachiewicz PL, Delley SS. The use of constained components in total hip arthroplasty. J Am Acad Orthop Surg 2002;10:4.
5
Kaper BP, Bernini PM, et al. Failure of a constrained acetabular prosthesis of a total hip arthroplasty:a report of four cases. J Bone Joint Surg 1998;80:561-5.
6
Thoms RJ, Marwin SE. A unique failure mechanism of a constrained total hip arthroplasty. J Arthroplasty 2008;23:293-8.
7
Tufescu TV, Dust W. Failure of a new constrained acetabular insert. J arthroplasty 2004;19:238-9.